5.1 Pointer
Fluctuation
Mechanical fluctuation of the
pointer is usually a symptom of excessive turbulence in the pipeline. This
is frequently caused by valves or fittings installed too close to the flow
meter inlet. if moving the flow meter is not practical, snubbers with
smaller holes can be installed to increase the mechanical damping of the
flow meter. (Pulsation snubbers are installed standard on all gas and liquid
flow meters unless ordered with viscosity correction).
5.2 Zero Shift
The need to
frequently re-zero the flow meter is an indication of excessive water hammer
or surge. Piping or operating procedures should be changed to reduce water
hammer in order to prolong the life of the flow meter.
5.3 Sluggish
Meter Response
Sluggish flow
meter response can be caused by inadequate pointer return spring preload or
by clogging of the pressure ports with debris. To check spring preload, see
"Checking the Pointer Preload".
To clean the pressure ports, see "CLEANING OF PRESSURE PORTS".
6.1 Liquid
Service
If the
specific gravity of the liquid being metered differs from the value marked
on the flow meter dial, a correction factor, J, must be applied to the flow
meter reading to obtain the actual flow rate. Click
here to access
calculator.
Actual
flow rate = J x flow meter reading
where:
Sc =
Specific Gravity for which the flow meter is calibrated
Sa = Actual
specific gravity of the fluid being metered.
Example:
Flow meter,
calibrated for water (specific gravity = 1.0) reads 12.5 GPM for a
lubricating oil with a specific gravity of 0.85. What is the actual flow
rate?

Therefore, actual flow = 1.085 x
12.5 = 13.56 GPM