4.3 Adjustment Pot
LocationPots VR1,
VR2 and VR3 are located just to the right of the terminal strip when viewing
the flow meter from the rear. VR1 is adjacent to terminal 10, VR2 is the
center pot and VR3 is the rightmost. See Figure 4.
4.4 Calibration
Note: The easiest way to set
the pointer for calibration of the transmitter is with the vertical lever.
By manipulating this lever, the pointer can be driven up/down scale to the
desired location., simulating the flow rate at any given point. (Take care
not to flex or push forward the lever as calibration will be affected).
Click
here for 4-20mA calibration video clip.
Adjust zero pot VR1 fully
CCW. Adjust 4mA output pot VR2 to 4.00mA. Set pointer to 30% using the lever
or flow. Adjust zero pot VR1 to 8.6mA. Set pointer to 95%. Adjust span pot
VR3 to 19.2mA. Repeat zero (VR1) and span (VR3) adjustment until both adjust
within specifications.
5.1 Field
Calibration
If the transmitter is fitted
with current output, adjust current output first. With the pointer at 95%
F.S. set VR6 pot to 950 Hz ± 30 Hz. If current output is not fitted, the
flow meter MUST be returned to the factory for calibration.
Important: When making
adjustments to VR3 take care not to push on the Hall effect sensor as this
will affect the calibration or cause damage.
Calibration is now complete
remove the multimeter, re-connect the wires to each peripheral and reinstall
the back cover plate.